Oktay Algın, Ergin Atalar and colleagues have published “Receive‐Only Coupled Wireless Radiofrequency Probe for Endocavity MR Imaging Using a pTx System” in Magnetic Resonance in Medicine (November 2026).
During interventional MRI, clinicians often need a clear view of one small region, for example tissue reached through a body cavity. The scanner’s built-in coils sit far from such regions and may not collect enough signal to show fine detail.
The study introduces a small wireless RF resonator probe that is placed close to the target and acts only as a receiver. The probe is made with standard microfabrication methods and tuned to the operating frequency of a 3T scanner. To keep it from interacting with the transmit field, the body birdcage coil of a parallel-transmit (pTx) system is driven in a linearly polarized mode. Simulations and temperature measurements show a normalized temperature rise below 0.1 °C, corresponding to a specific absorption rate under 1.2 W/kg. The probe raises the signal-to-noise ratio 1.6-fold in phantom experiments and about two-fold near the probe in vivo.
The approach adds image quality exactly where it is needed while keeping heating low. It could make interventional MRI safer and help clinicians see small target regions more clearly.
Authors: Akbar Alipour, Volkan Açıkel, Sayım Gökyar, Oktay Algın, Çağdaş Oto, Priti Balchandani, Hilmi Volkan Demir, Ergin Atalar